CLEANING

19

Every day the caustic pump, delivery system and distribution head should be cleaned as described
in the ‘ROUTINE MAINTENANCE’ section of this manual.

Accumulated deposits or scale in the boiler may be removed by removing the boiler and washing it
with a dilute solution of hydrochloric acid or a commercially available scale remover that is
compatible with glass.

USE APPROPRIATE PROTECTIVE SAFETY EQUIPMENT WHEN
HANDLING CLEANING SOLUTIONS. ALWAYS FOLLOW THE
CLEANING SOLUTION INSTRUCTIONS REGARDING THEIR USE,
SAFETY PRECAUTIONS TO BE FOLLOWED AND DISPOSAL.


For instruction on how to remove the boiler, refer to the ‘Removing the Boiler’ in the ‘SERVICE’
section of this manual.

The exterior of the unit should also be cleaned after each work session using a soft cloth and a
solution of mild soap and water to prevent the accumulation of deposits.