Refer to the following table if the system fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco for
additional assistance.
PROBLEM
CAUSE
CORRECTIVE ACTION
Blower and controller
won’t turn on
Unit not plugged into
outlet
Plug the cabinet into appropriate electrical
service.
Circuit breaker tripped
Reset circuit breaker.
Prints are not
developing properly
Insufficient humidity in
the clear chamber
Place 1-2 drops of water in a heating pan,
and place on the heater.
Little or no fumes are
being produced
Heater temperature is
too low.
Increase heater temperature
Bottom of heater pan is
dented/bent
Discard damaged pan, and use one with flat
bottom.
Heater temperature not
changing on controller
(“out” indicator is on)
Heater /Over
temperature switch
failure
Contact Labconco Product Service
Department.
Heater temperature not
changing on controller
(out indicator is off)
Controller is defective
Contact Labconco Product Service
Department.
Fan does not run
Fan or 12 volt power
supply failure
Contact Labconco Product Service
Department.
Controller displays “Er”
or “oo” or “-”
Controller error
Contact Labconco Product Service
Department.