Model Series
36000
36125
36100
Width (feet)
3
4
6
HEPA Filter Dims. (in.)
36 x 30 x 6
48 x 30 x 6
72 x 30 x 6
Average Velocity Reading Grid (inches)
12 x 12
12 x 12
12 x 12
Grid Distance from Diffuser (inches)
6
6
6
Grid Distance from Sidewalls (inches)
6
6
6
Average Velocity (FPM)
90 ±10
90 ±10
90 ±10
Work Area (ft
2
) 7.5
10
15
Exhaust Volume Range (CFM)
600-750
800-1,000
1,200-1,500
Number of Laskin Nozzles needed
1
2
2
APPROXIMATE SETTINGS FOR NEW FILTERS
Mag Gauge Reading
0.7 – 1.0
0.7 – 1.0
0.7 – 1.0
Motor HP
1/3
1/2
2 1/3
Speed Winding used
High
High
High
Motor Capacitator
15
15
15
Fluorescent Light(s)
F25T8/TL741 F32T8/TL741
F72T12CW
UV Light(s) (Option)
G30T8
G30T8
G64T5L/CB