Service operations – Labconco Purifie Horizontal Clean Benches 36125 Series User Manual
Page 48
![background image](/manuals/738631/48/background.png)
44
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Clean Bench
Weekly
• Wipe down the interior surfaces of the clean
bench with a mild household detergent or
cleaner and allow to dry.
• Using a damp cloth, clean the exterior surfaces
of the clean bench, particularly the front and top
of the clean bench to remove any accumulated
dust.
Monthly (or more often as required)
• Check all service valves, if so equipped, for
proper operation.
• Check the prefilters and replace if necessary.
The prefilters should be replaced at least
quarterly.
• All weekly activities.
Quarterly
• Replace the prefilter. See page 57 for ordering
information.
• All monthly activities.
Annually
• Have the clean bench recertified by a qualified
certification technician.
• All quarterly activities.
Biannually
• Replace the fluorescent lamp if required.
• Replace the UV lamp (if included).
• All annual activities.
Service Operations
Resetting the Circuit Breaker:
1. The circuit breaker is located next to the power cord
on the upper rear panel of the bench as shown in
Figure 6-1. If the circuit breaker trips, it can be
reset by pressing the white button in.