Initial certification – Labconco Purifie Horizontal Clean Benches 36125 Series User Manual
Page 24

20
Product Service 1-800-522-7658
Chapter 3: Getting Started
Initial Certification
Prior to use, all Purifier Clean Benches should be
certified by a qualified certifier. Under normal
operating conditions, the Purifier Clean Bench should
be recertified at least annually, and when moved or
serviced. The certifier should perform the following
tests, as recommended in Institute of Environmental
Sciences and Technology, IEST RP-CC0002.2:
• Airflow Velocity Test
• HEPA Filter Leak Test
• Introduction Leak Test/Backstreaming
(when appropriate)
• Airborne Particle Count (when
appropriate)
• Lighting Intensity Test (when
appropriate)
• Noise Level Test (when appropriate)
• Vibration Test (when appropriate)
Figure 3-4