Labconco Biological Safety Cabinets 34610 Series User Manual
Page 75
Product Service 1-800-522-7658
70
A
A
p
p
p
p
e
e
n
n
d
d
i
i
x
x
E
E
:
:
Q
Q
u
u
i
i
c
c
k
k
C
C
h
h
a
a
r
r
t
t
Model (Series)
34410xx
34610xx
Cabinet Size (in feet)
4
6
Sash Opening (inches)
8
8
Starting Serial #
1
0705_
0705_
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
Nominal Average Inflow (FPM)
105+/-5
105+/-5
Nominal Avg. Inflow Vol. (CFM)
283 +/-13
423+/-20
DIM Exhaust Volume (CFM)
665
998
Traverse Exhaust Volume (CFM)
810
1218
Differential Pressure (in. H
2
0) 1.9
2.2
Supply HEPA Data
Labconco P/N
3838401
3838403
Exhaust HEPA Data
Labconco P/N
3838501
3838503
Motor/Blower Data
Labconco P/N
2
3832204
3832205
Motor HP
½
½
Blower Size (inches)
10 x 8
10 x 8
Fluorescent/UV Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F40T8 SP41
Light intensity (ft-candles)
3
90-100 90-100
UV Lamp
G30T8
G30T8
UV energy(uW-cm
2
)
4
240
240
B2 Prefilter Data
Labconco P/N
3850500
3850501
1. The primary serial tag is on the lower outside edge of the right corner
post. The secondary serial tag is located on the front of the electronics
module on the top right side of the cabinet. The first two digits of the
serial number are the year of production, the next two are the month.
The next 5 digits are the sequence of production, and the letter
following the serial number is the revision level of the cabinet.
2. Each motor must be programmed by Labconco for the appropriate
width cabinet.
3. As per ANSI/NSF Std. 49 with a background of 15 ft-candles.
4. Measured in the geometric center of the work surface.