Chapter 6: Maintaining The Cabinet
Product Service 1-800-522-7658
51
4. Reinstall the grille by reversing the above sequence, ensuring that the
grille properly engages the bottom edge of the cabinet.
Towel Catch Removal:
Although not normally required, the towel catch can be removed for cleaning,
inspection, etc.
Note: The work surface of the cabinet and the towel catch must be
thoroughly decontaminated before removing either.
1. Remove the work surface as described above.
2. Lift the front end of the center support straight up to release it from the
locating pins (A). Set the support aside in the work area.
3. Remove the towel catch by pivoting the bottom out toward you (B).
4. Surface decontaminate the towel catch before removing it.
5. Reinstall the towel catch by sliding it back into position, ensuring that the
two slots on either end of the towel catch properly engage the rails on the
rear of the duct, as shown in Figure 6-2. Also ensure that the DOP
Sampling Line (the clear tube with a black cap) either passes through a
hole in the towel catch, or passes along the outside edge of the catch
without being kinked.