Hanna Instruments HI 98186 User Manual
Page 58

58
•
maximum time in seconds (4 chars)
•
minimum start DO including sign and decimal point (6 chars)
•
minimum end DO including sign and decimal point (6 chars)
•
total volume including sign and decimal point (6 chars)
•
sample volume including sign and decimal point (6 chars)
•
solids weight including sign and decimal point (6 chars)
•
SOUR temperature correction (1=enabled, 0=disabled)
(1char)
•
The short name of the selected language (3 chars)
NSLx Requests the number of logged samples (4 chars)
The command parameter (x - 1 char)
•
D - - - - - the request is for DO
•
B -
-
-
-
- the request is for BOD
•
O -
-
-
-
- the request is for OUR
•
S - the request is for SOUR
•
I - the request is for BOD initial values
LODDxxx Requests the xxx
th
DO logged data
LODBxxx Requests the xxx
th
BOD logged data
LODOxxx Requests the xxx
th
OUR logged data
LODSxxx Requests the xxx
th
SOUR logged data
LODIxxx
Requests the xxx
th
initial BOD value logged data
LODDALL Requests all the DO log on demand
LODBALL Requests all the BOD log on demand
LODOALL Requests all the OUR log on demand
LODSALL Requests all the SOUR log on demand
LODIALL Requests all the initial BOD values log
The answer string for each record contains:
•
The logged mode(2 chars)
•
20 - DO range
•
21 - BOD range
•
22 - OUR range
•
23 - SOUR range
•
24 - BOD initial values
•
DO log data:::::
•
Measurement unit (0 = %, 1 = mg/L) (1 char)
•
DO value including sign and decimal point (8 chars)
•
Salinity value [g/L] (3 chars)
•
Pressure value in mmHg, including sign and decimal point
(11 chars)
•
Temperature value including sign and decimal point
(8 chars)