buffer value including sign and decimal point (this value is sent only
if there is a 2 point calibration) (6 chars)
•
salinity value (3 chars)
•
pressure value in mmHg, including sign and decimal point
(11 chars)
•
temperature value including sign and decimal point (8 chars)
•
calibration time: yymmddhhmmss (12 chars).
PAR Requests the setup parameters setting.
The answer string contains:
•
backlight value (1 char)
•
contrast value (2 chars)
•
instrument ID (4 chars)
•
calibration alarm timeout (2 chars)
•
Setup information (2 chars): 8 bit hexadecimal encoding:
•
0x01: -
-
-
-
- beep is on
•
0x04: -
-
-
-
- Celsius / Fahrenheit display (°C if the bit is set)
•
0x10: - manual pressure (1 if activated, 0 otherwise)
•
Auto Light-off time (3 chars)
•
Auto Power-off time (3 chars)
•
Salinity value (3 chars)
•
Pressure unit (1 char): 0 - mmHg, 1 - inHg, 2 - atm, 3 - mbar,
4 - psi, 5 - kPa
•
BOD configuration values
•
sample minimum delta DO, including sign and decimal point
(6 chars)
•
sample minimum end DO, including sign and decimal point
(6 chars)
•
seed minimum delta DO, including sign and decimal point
(6 chars)
•
seed minimum end DO, including sign and decimal point
(6 chars)
•
OUR configuration values
•
minimum time in seconds (4 chars)
•
maximum time in seconds (4 chars)
•
minimum start DO including sign and decimal point (6 chars)
•
minimum end DO including sign and decimal point (6 chars)
•
total volume including sign and decimal point (6 chars)
•
sample volume including sign and decimal point (6 chars)
•
SOUR configuration values
•
minimum time in seconds (4 chars)