BOD configuration
Focus on
BOD configuration item.
Press Select.
Parameters:
•
Sample min
D
DO -the minimum acceptable difference between the
initial and final DO values for a sample. If the difference is less than this
value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
•
Sample min end DO - the minimum acceptable final DO value for a
sample. If the final DO value is less than this value the meter will show a
warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
•
Seed min
D
DO - the minimum acceptable difference between the initial
and final DO values for a seed sample. If the difference is less than this
value the meter will show a warning message when evaluating the BOD.
Range: 0.00 to 50.00 mg/L.
•
Seed min end DO - the minimum acceptable final DO value is less than
this value the meter will show a warning message when evaluating the
BOD.
Range: 0.00 to 50.00 mg/L.
Press Prev/Next to select a different parameter.
Press ARROW keys to modify the selected parameter’s value.
Press Save to save the new BOD configuration.
Press ESC to leave without changing.