Hanna Instruments HI 98186 User Manual
Page 37

37
OUR configuration
Focus on
OUR configuration item.
Press Select.
Parameters:
•
Min time - the minimum time for the OUR test.
Range: 1 to 3600 seconds.
•
Max time - the maximum time for the OUR test. The test will stop
automatically when the maximum time has elapsed.
•
Min start
DO - the minimum accepted DO value for starting the OUR test.
If the DO reading is less than this value the test cannot be started.
Range: 0.00 to 50.00 mg/L.
•
Min end DO - the minimum accepted DO value at the end of the test. If
the DO reading at the end of the OUR test is less than this value a
warning message will be displayed.
Range: 0.01 to 50.00 mg/L..
•
Total volume - the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
•
Sample volume - the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
Press Prev/Next to select a different parameter.
Press ARROW keys to modify the selected parameter’s value.
Press Save to save the new OUR configuration.
Press ESC to leave without changing.
SOUR configuration
Focus on
SOUR configuration item.