beautypg.com

Hanna Instruments HI 83203-2008 User Manual

Page 6

background image

10

11

POWER CONNECTION AND BATTERY MANAGEMENT

The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.
Note: Always turn the meter off before unplugging it to ensure no data is lost.
When the meter switches ON, it verifies if the power supply adapter is connected. The battery icon on the
LCD will indicate the battery status:
- battery is charging from external adapter

- battery fully charged (meter connected to AC/DC adapter)

- battery capacity (no external adapter)

- battery Low (no external adapter)

- battery Dead (no external adapter)

OPERATIONAL GUIDE

• Turn ON the instrument via the ON/OFF power switch.
• The meter will perform an autodiagnostic test. During this test, the Hanna Instrument logo will appear on

the LCD. After 5 seconds, if the test was successful, the last selected method will appear on the display.

• In order to select the desired method press METHOD and a screen with the available methods will appear.
• Press keys to highlight the desired method. Press Select.

METHOD SELECTION

METHOD REFERENCE TABLE

HI 83203 - AQUACULTURE

Method

Method

Page

description

1

Ammonia MR

16

2

Ammonia LR

18

3

Free Chlorine

20

4

Total Chlorine

23

5

Copper HR

26

6

Copper LR

28

7

Nitrate

30

HEALTH & SAFETY

The chemicals contained in the reagent kits may be hazardous if improperly handled. Read the
Material Safety Data Sheet (MSDS) before performing tests.

Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions carefully.
Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for

each sample and to use the same cuvette for zeroing and measurement when possible (for most precise
results follow the measurement procedures carefully).

• Discard the sample immediately after the reading is taken, or the glass might become permanently stained.
• All the reaction times reported in this manual are at 20°C (68°F). As a general rule of thumb, they

should be doubled at 10°C (50°F) and halved at 30°C (86°F).

REAGENT BLANK CORRECTION

• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly

in the same way, only the blank is deionized water instead of sample.
A blank cuvette may be used more than once: stability and storing conditions are described for each
method in the related chapter.

INTERFERENCES

• In the method measurement section the most common interferences that may be present in an average

wastewater matrix have been reported. It may be that for a particular treatment process other
compounds do interfere with the method of analysis.

Method

Method

Page

description

8

Nitrite HR

32

9

Nitrite LR

34

10

Dissolved Oxygen

36

11

pH

38

12

Phosphate HR

40

13

Phosphate LR

42