Hanna Instruments HI 83203-2008 User Manual
Page 12

22
23
Total Chlorine
TOTAL CHLORINE
SPECIFICATIONS
Range
0.00 to 3.50 mg/L
Resolution
0.01 mg/L
Accuracy ±0.03 mg/L ±3% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
EPA DPD method 330.5. The reaction between the chlorine and the
DPD reagent causes a pink tint in the sample.
REQUIRED REAGENTS
POWDER:
Code
Description
Quantity
HI 93711-0
DPD
1 packet
LIQUID:
Code
Description
Quantity
HI 93701A-T
DPD1 indicator
3 drops
HI 93701B-T
DPD1 buffer
3 drops
HI 93701C
DPD3 solution
1 drop
REAGENT SETS
HI 93701-T Reagents for 300 total chlorine tests (liquid)
HI 93711-01 Reagents for 100 total chlorine tests (powder)
HI 93711-03 Reagents for 300 total chlorine tests (powder)
For other accessories see page 46.
MEASUREMENT PROCEDURE
• Select the
Total Chlorine method using the proce-
dure described in the
Method Selection section (see
page 11).
• Fill the cuvette with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
10 mL
• Reinsert the cuvette into the instrument.
• Press READ to start the reading. The instrument displays the results
in mg/L of free chlorine.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately 2
minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Free Chlorine