beautypg.com

Hanna Instruments HI 83203-2008 User Manual

Page 21

background image

40

41

Phosphate HR

• Add the content of one packet of HI 93717B-0 Phosphate

HR Reagent B to the cuvette. Replace the cap and shake
gently until completely dissolved.

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior

to the measurement or, alternatively, wait for 5 minutes
and press READ. When the timer ends the meter will
perform the reading. The instrument displays the results in
mg/L of phosphate (PO

4

3-

).

• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus

pentoxide (P

2

O

5

).

• Press the or to go back to the measurement screen.

INTERFERENCES

Sulfide
Chloride above 150000 mg/L )
Calcium above 10000 mg/L as CaCO

3

Magnesium above 40000 mg/L as CaCO

3

Ferrous iron above 100 mg/L

Phosphate HR

PHOSPHATE HIGH RANGE

SPECIFICATIONS

Range

0.0 to 30.0 mg/L

Resolution

0.1 mg/L

Accuracy

±1 mg/L ±4% of reading

Typical EMC Dev. ±0.1 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Amino Acid method. The reaction between phosphate and reagents

causes a blue tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93717A-0

Molybdate

10 drops

HI 93717B-0

Reagent B

1 packet

REAGENT SETS

HI 93717-01 Reagents for 100 tests
HI 93717-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Phosphate HR method using the procedure de-

scribed in the

Method Selection section (see page 11).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when the meter

is zeroed and ready for measurement.

• Remove the cuvette.

• Add 10 drops of HI 93717A-0 Molybdate reagent.

10 mL

x 10