Phosphate LR

• Reinsert the cuvette into the instrument.

• Press TIMER and the display will show the countdown prior to the

measurement or, alternatively, wait for 3 minutes and press READ.
When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of phosphate (PO

4

3-

).

• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of phosphorus (P) and phosphorus

pentoxide (P

2

O

5

).

• Press the or to go back to the measurement screen.

INTERFERENCES

Interference may be caused by:
Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered samples also interfere.

Phosphate LR

PHOSPHATE LOW RANGE

SPECIFICATIONS

Range

0.00 to 2.50 mg/L

Resolution

0.01 mg/L

Accuracy

±0.04 mg/L ±4% of reading

Typical EMC Dev. ±0.01 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Ascorbic Acid method. The reaction between phosphate and the

reagent causes a blue tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93713-0

Powder reagent

1 packet

REAGENT SETS

HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 46.

MEASUREMENT PROCEDURE

• Select the

Phosphate LR method using the procedure

described in the

Method Selection section (see page 11).

• Rinse, cap and shake the cuvette several times with unreacted

sample. Fill the cuvette with 10 mL of sample up to the
mark and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press ZERO key. The display will show “-0.0-” when the meter is

zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of

HI 93713-0 reagent. Replace the cap and shake gently
(for about 2 minutes) until the powder is completely
dissolved.

10 mL