Silica
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when
the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI 93705C-0 Amino acid reagent and shake until it is
completely dissolved.
• Reinsert the cuvette into the instrument.
• Press Timer and the display will show the countdown prior to the measurement. Alternatively, wait for
exactly 3 minutes and press Read. When the timer ends the meter will perform the reading. The
instrument displays concentration in mg/L of silica (SiO
2
).
• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of Silicon (Si).
• Press or to return to the measurement screen.
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)