beautypg.com

Hanna Instruments HI 83205 User Manual

Page 37

background image

37

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 6 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in μμμμμg/L of chromium VI.

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in μg/L of Chromate (CrO

4

2

¯) and Dichromate (Cr

2

O

7

2

¯).

• Press or to return to the measurement screen.

INTERFERENCES

Interference may be caused by:

Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed.

Iron above 1 ppm

Mercurous and mercuric ions cause slight inhibition of the reaction.

Chromium VI LR