beautypg.com

Hanna Instruments HI 83205 User Manual

Page 22

background image

22

Ammonia LR

• Remove the cuvette.

• Add 4 drops of HI 93700A-0 First reagent (6 drops for

seawater analysis). Replace the cap and mix the solution.

• Add 4 drops of HI 93700B-0 Second reagent (10 drops for

seawater analysis). Replace the cap and mix the solution.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the

measurement or, alternatively, wait for 3 minutes and 30 seconds and
press Read. When the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of ammonia nitrogen
(NH

3

-N).

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of ammonia (NH

3

) and ammonium (NH

4

+

).

• Press or to return to the measurement screen.

INTERFERENCES

Interference may be caused by: acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic

chloramines, sulfide, various aliphatic and aromatic amines.