• Add 12 drops of the HI 93704-0 reagent to

each cuvette. Replace the caps and shake
gently to mix (about 30 seconds).

• Place the blank (#1) into the holder and close the lid.

• Press Timer and the display will show the countdown prior to zeroing the blank. The display will show

“-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank.
• Insert the cuvette with the reacted sample (# 2) into the

instrument and close the lid.

• Press Read to start the reading. The instrument displays concentration in μμμμμg/L of hydrazine.

Hydrazine

INTERFERENCES

Interference may be caused by:
Highly colored samples
Highly turbid samples
Aromatic amines