Hanna Instruments HI 83206 User Manual

Page 42

background image

42

Nickel HR

SPECIFICATIONS
Range

0.00 to 7.00 g/L

Resolution

0.01 g/L

Accuracy

±0.07 g/L ±4% of reading at 25 °C

Typical EMC

±0.02 g/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the photometric method. The reaction between nickel and the reagent
causes a blue tint in the sample.

REQUIRED REAGENTS
Code

Description

Quantity

HI 93726-0

Powder reagent 1 packet

REAGENT SETS
HI 93726-01 Reagents for 100 tests
HI 93726-03 Reagents for 300 tests
For other accessories see page 73.
MEASUREMENT PROCEDURE
• Select the

Nickel HR method using the procedure described in the Method Selection section (see page 12).

• Fill the cuvette up to the mark with 10 mL of unreacted sample

and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

NICKEL HIGH RANGE

10 mL