Molybdenum

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays concentration in mg/L of molybdenum.

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of molybdate (MoO

4

) and sodium molybdate

(Na

2

MoO

4

).

• Press or to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Aluminum above 50 mg/L
Chromium above 1000 mg/L
Copper above 10 mg/L
Iron above 50 mg/L
Nickel above 50 mg/L
Nitrite, as NO

2

¯

Sulfate above 200 mg/L
Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.