beautypg.com

Hanna Instruments HI 83206 User Manual

Page 10

background image

10

• Pay attention to push the cuvette completely down in the holder and to align the white point on the

cap to the indexing mark on the meter.

• In order to avoid reagent leaking and to obtain more accurate

measurements, close the cuvette first with the supplied HDPE plastic
stopper

and then the black cap.

• Whenever the cuvette is placed into the measurement cell, it must be dry

outside, and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing higher

readings. To obtain accurate measurements, remove such bubbles by
swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added. For

best accuracy, respect the timings described in each specific method.

• It is possible to take multiple readings in a row, but it is recommended to

take a new zero reading for each sample and to use the same cuvette for
zeroing and measurement when possible (for most precise results follow the
measurement procedures carefully).

• Discard the sample immediately after the reading is taken, or the glass

might become permanently stained.

• All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time

should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher
than 25 °C (77 °F).

INTERFERENCES

• In the method measurement section the most common interferences that may be present in an average

sample matrix have been reported. It may be that for a particular treatment process other compounds
do interfere with the method of analysis.