Cyanuric Acid

CYANURIC ACID

SPECIFICATIONS

Range

0 to 80 mg/L

Resolution

1 mg/L

Accuracy

±1 mg/L ±15% of reading at 25 °C

Typical EMC

±1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the turbidimetric method. The reaction between cyanuric acid and the
reagent causes a white suspension in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93722-0

Powder reagent

1 packet

REAGENT SETS

HI 93722-01 Reagents for 100 tests
HI 93722-03 Reagents for 300 tests
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Cyanuric Acid method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Add the content of one packet of HI 93722-0 Cyanuric

Acid Reagent. Replace the cap and shake gently for
about 10 seconds (dissolution is not complete).

10 mL