beautypg.com

Hanna Instruments HI 83206-2008 User Manual

Page 21

background image

40

41

Molybdenum

• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press READ. When the timer ends the meter will perform the reading. The instrument

displays concentration in mg/L of molybdenum.

• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in mg/L of molybdate (MoO

4

) and sodium

molybdate (Na

2

MoO

4

).

• Press the or to go back to the measurement screen.

INTERFERENCES

Interference may be caused by:

Aluminum above 50 mg/L

Chromium above 1000 mg/L

Copper above 10 mg/L

Iron above 50 mg/L

Nickel above 50 mg/L

Nitrite, as NO

2

Sulfate above 200 mg/L

Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents.

• Fill one graduated mixing cylinder up to the 25 mL

mark with the sample.

• Add the content of one packet of HI 93730A-0

molybdenum reagent, close the cylinder and invert it

several times until completely dissolved.

• Add the content of one packet of HI 93730B-0

molybdenum reagent to the cylinder, close and invert

it several times until completely dissolved.

• Add the content of one packet of HI 93730C-0

molybdenum reagent to the cylinder, close and shake

it vigorously.

• Fill an empty cuvette with 10 mL of sample up to the

mark and replace the cap.

• Insert the cuvette into the instrument.

Molybdenum