Color of Water

COLOR OF WATER

SPECIFICATIONS

Range

0 to 500 PCU (Platinum Cobalt Units)

Resolution

1 PCU

Accuracy

±10 PCU ±5% of reading

Typical EMC

± 1 PCU

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Colorimetric Platinum Cobalt method.

REQUIRED ACCESSORIES

0.45 µm membrane for true color measurement.
For other accessories see page 73.

MEASUREMENT PROCEDURE

• Select the

Color of Water method using the procedure described in

the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with deionized water and replace

the cap. This is the blank.

• Place the blank (# 1) into the holder and close the lid.

• Press ZERO key. The meter will show “-0.0-” when the meter is

zeroed and ready for measurement.

• Remove the blank.

10 mL

# 1

# 1

• Reinsert the cuvette into the instrument.
• Press TIMER and the display will show the countdown prior to the

measurement or, alternatively, wait for 6 minutes and press

READ. When the timer ends the meter will perform the reading.
The instrument displays concentration in µg/L of chromium VI.

• Press the or to access the second level of functions.
• Press the Chem Frm functional key to convert the result in µg/L of Chromate (CrO

4

2-

) and Dichromate

(Cr

2

O

7

2-

).

• Press the or to go back to the measurement screen.

INTERFERENCES

Interference may be caused by:

Vanadium above 1 ppm. However, waiting 10 minutes before reading, the interference is removed.

Iron above 1 ppm

Mercurous and mercuric ions cause slight inhibition of the reaction.

Chromium VI LR