Silver

SILVER

SPECIFICATIONS

Range

0.000 to 1.000 mg/L

Resolution

0.005 mg/L

Accuracy

±0.020 mg/L ±5% of reading at 25 °C

Typical EMC

± 0.001 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm.

Method

Adaptation of the PAN method. The reaction between silver and reagents causes an
orange tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93737A-0

Buffer Reagent A

1 mL

HI 93737B-0

Buffer Reagent B

1 mL

HI 93737C-0

Indicator Reagent C 2 mL

HI 93737D-0

Fixing Reagent D

2 mL

HI 93703-51

Dispersing Agent

4-6 drops

REAGENT SETS

HI 93737-01 Reagents for 50 tests
HI 93737-03 Reagents for 150 tests
For other accessories see page 49.

MEASUREMENT PROCEDURE

• Select the

Silver method using the procedure described in the Method Selection section (see page 12).

Note: for best results perform your tests between 20-24°C.

• Fill two graduated beakers with 25 mL of sample.

• Add 1.0 mL of HI 93737A-0 Buffer reagent to one beaker

(the blank) and swirl gently to mix.

# 1

# 2

# 1