Chlorine Dioxide

CHLORINE DIOXIDE

SPECIFICATIONS

Range

0.00 to 2.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.10 mg/L ±5% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and
reagents causes a colorless to purple tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93738A-0

Reagent A

1 mL

HI 93738B-0

Dechlorinating Reagent B 1 packet

HI 93738C-0

Reagent C

1 mL

HI 93738D-0

Reagent D

1 mL

REAGENT SETS

HI 93738-01 Reagents for 100 tests

HI 93738-03 Reagents for 300 tests

For other accessories see page 49.

MEASUREMENT PROCEDURE

• Select the

Chlorine Dioxide method using the procedure

described in the

Method Selection section (see page 12).

• Fill two graduated mixing cylinders (#1 & #2) up to the

25 mL mark with the sample.

• Add 0.5 mL of HI 93738A-0 Chlorine Dioxide Reagent to each cylinder (#1 & #2), close them and

invert several times to mix.

# 2

25 mL

# 1

25 mL

# 1 & # 2

# 1

# 2