Free Chlorine

FREE CHLORINE

SPECIFICATIONS

Range

0.00 to 2.50 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

EPA DPD method 330.5. The reaction between free chlorine and the

DPD reagent causes a pink tint in the sample.

REQUIRED REAGENTS

POWDER

:

Code

Description

Quantity

HI 93701-0

DPD

1 packet

LIQUID:
Code

Description

Quantity

HI 93701A-F

DPD1 Indicator

3 drops

HI 93701B-F

DPD1 Buffer

3 drops

REAGENT SETS

HI 93701-F Reagents for 300 tests (liquid)
HI 93701-01 Reagents for 100 tests (powder)
HI 93701-03 Reagents for 300 tests (powder)
For other accessories see page 49.

MEASUREMENT PROCEDURE

• Select the

Free Chlorine method using the procedure described

in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

10 mL