Hanna Instruments HI 83215-2008 Series User Manual

Page 42

background image

42

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette and add the content of one packet of

HI 93750B-0 reagent. Replace the cap and gently mix for
one minute by slowly turning the cuvette upside down.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 2 minutes and press Read. When the timer ends the meter will perform the reading. The instrument

displays the results in mg/L (ppm) of potasium (K).

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium oxide (K

2

O)

• Press or to return to the measurement screen.
• For ULTRA HIGH RANGE samples: follow the procedure described at page 43.

Potassium HR