Hanna Instruments HI 83215-2008 Series User Manual
Page 23
![background image](/manuals/737581/23/background.png)
23
Ammonia HR
SPECIFICATIONS
Range
0 to 100 mg/L
Resolution
1 mg/L
Accuracy
±1 mg/L ±4% of reading at 25 °C
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 420 nm
Method
Adaptation of the
ASTM Manual of Water and Environmental Technology, D1426-92,
Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93715A-0
First Reagent
4 drops (6 drops for seawater)
HI 93715B-0
Second Reagent 4 drops (10 drops for seawater)
REAGENT SETS
HI 93715-01 Reagents for 100 tests
HI 93715-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (HR)
procedure at page 21.
• Select the
Ammonia HR method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
AMMONIA HIGH RANGE
10 mL