Hanna Instruments HI 83215-2008 Series User Manual
Page 29
29
SPECIFICATIONS
Range
0 to 300 mg/L
Resolution
5 mg/L
Accuracy
±10 mg/L ±8% of reading at 25 °C
Typical EMC
±5 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the cadmium reduction method. The reaction between nitrate-nitrogen
and the reagent causes an amber tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93728-0
Powder reagent 1 packet
REAGENT SETS
HI 93728-01 Reagents for 100 tests
HI 93728-03 Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Note: for sample preparation follow the NUTRIENTS SOLUTION (HR) procedure at page 21.
• Select the
Nitrate HR method using the procedure
described in the
Method Selection section (see page 12).
• Using the pipette, fill the cuvette with 6 ml of sample,
up to half of its height, and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette and add the content of one packet of
HI 93728-0 reagent.
NITRATE HIGH RANGE
Nitrate HR
6 mL