The instrument shows clear warning messages when erroneous conditions appear and when measured values are
outside the expected range. These messages are described bellow.

No Light: The light source is not functioning properly.

Light Leak: There is an excess amount of ambient light reaching the
detector.

Inverted cuvettes: The sample and the zero cuvettes are inverted.

Battery Low: The battery capacity is lower than 10%.

Light Low: The instrument cannot adjust the light level. Please check
that the sample does not contain any debris.

ERRORS AND WARNINGS

Light High: There is too much light to perform a measurement. Please
check the preparation of the zero cuvette.

• Remove the cuvette and add 5 drops of HI 93710-0

Phenol Red Indicator. Replace the cap and mix the
solution.

• Reinsert the cuvette into the instrument.

• Press the

Read key to start the reading. The instrument displays the pH value.