Hanna Instruments HI 83216 User Manual
Page 16

30
31
pH
pH
SPECIFICATIONS
Range
6.5 to 8.5 pH
Resolution
0.1 pH
Accuracy
±0.1 pH at 25 °C
Typical EMC
±0.1 pH
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93710-0
Phenol Red Indicator
5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
For other accessories see page 34.
MEASUREMENT PROCEDURE
• Select the
pH method using the procedure described in the Method Selection section (see page 15).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Add the content of one packet of HI 93722-0 Cyanuric Acid
Reagent. Replace the cap and shake gently for about 10
seconds (dissolution is not complete).
• Reinsert the cuvette into the instrument.
• Press
Timer and the display will show the countdown prior to the measurement or, alternatively, wait
for 45 seconds and press
Read. When the timer ends the meter will perform the reading.
• The instrument displays concentration in
mg/L of cyanuric acid.
INTERFERENCES
Turbidity preexisting in the sample causes interference during measurement.
Cyanuric Acid
10 mL