SPECIFICATIONS

Range

0 to 100 mg/L

Resolution

5 mg/L

Accuracy

±5 mg/L ±3% of reading at 25 °C

Light Source

Tungsten lamp with narrow band interference filter @ 466 nm

Method

Sulfate is precipitated with barium chloride crystals. Light absorbance of the suspension

is measured.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93751-0

Indicator reagent

1 packet

REAGENT SETS

HI 93751-01 Reagents for 100 tests
HI 93751-03 Reagents for 300 tests
For other accessories see page 57.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the IRRIGATION WATER (LR) procedure on page 20.

• Select the

Sulfate method using the procedure described

in the

Method Selection section (see page 12).

• Fill a cuvette with 10 mL of unreacted sample (up to

the mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

Sulfate

SULFATE

10 mL