beautypg.com

Hanna Instruments HI 83225-01 User Manual

Page 42

background image

42

Potassium HR

POTASSIUM HIGH RANGE

SPECIFICATIONS

Range

20 to 200 mg/L

Resolution

5 mg/L

Accuracy

±30 mg/L ±7% of reading at 25 °C

Typical EMC Dev. ±5 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93750A-0

Potassium Reagent

6 drops

HI 93750B-0

Powder Reagent

1 packet

REAGENT SETS

HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 57.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the NUTRIENT SOLUTIONS (HR) procedure at page 22.
• Select the

Potassium HR method using the procedure described in the Method Selection section (see

page 12).

• Fill the cuvette with 10 mL of sample, up to the mark.

• Add six drops of HI 93750A-0, replace the cap and

swirl the solution.

• Place the cuvette into the holder and close the lid.

10 mL