29

Ammonia LR

• Remove the cuvette.

• Add 4 drops of HI 93715A-0 First Reagent (6 drops for seawater

analysis). Replace the cap and mix the solution by inverting the
cuvette a couple of times.

• Add 4 drops of HI 93715B-0 Second Reagent (10 drops for seawater

analysis). Replace the cap and mix the solution by inverting the
cuvette a couple of times.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the

measurement or, alternatively, wait for 3 minutes and 30 seconds
and press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L of ammonia
nitrogen (NH

3

-N).

• Press

or

to access the second level functions.

• Press the Chem Frm key to convert the result in mg/L of ammonia (NH

3

) and ammonium (NH

4

+

).

• Press

or

to return to the measurement screen.

INTERFERENCES

Interference may be caused by: acetone, alcohols, aldehydes, glycine, hardness above 1 g/L, iron, organic

chloramines, sulfide, various aliphatic and aromatic amines.