Hanna Instruments HI 83225-01 User Manual

Page 43

background image

43

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet

of HI 93750B-0 reagent. Replace the cap and gently
mix for one minute by slowly turning the cuvette upside
down.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 2 minutes and press Read. When the timer ends the meter will perform the reading. The instrument
displays the results in mg/L (ppm) of potassium (K).

• Press

or

to access the second level functions.

• Press the Chem Frm key to convert the result in mg/L of potassium oxide (K

2

O).

• Press

or

to return to the measurement screen.

• For ULTRA HIGH RANGE samples: follow the procedure described at page 44.

Potassium HR