Hanna Instruments HI 83749 User Manual
Page 19

19
• Replace the cap. Invert several times to mix and
then wait for 1 minute.
• Insert the cuvet into the instrument and close
the lid.
• Press READ and “----” will blink on the
display. After a few seconds the instrument
displays the first reading. The meter continues
updating the readings until the “Lamp and Read
Status” indicator turns off.
• The final displayed value is the averaged reading
of turbidity in NTU.
This is T2. Record the value.
• If “T2 stable. Otherwise the wine needs to be stabilized. Note: to get more representative results of long term protein stability, HANNA Instruments ® recommends to filter the wine sample first through a 0.45 micron filterdisc before analysis. e.g. T2 22 • Place the cuvet #1 into the instrument and close the lid. • Press READ and “----” will blink on the display. After a few seconds the instrument displays the first reading. The meter continues updating the readings until the “Lamp and Read Status” indicator turns off. • The final displayed value is the averaged reading. This is T1 for sample #1. Record the value. • Remove cuvet #1 from the instrument and open the cap. • Use the 1000 µL automatic pipette to add exactly 1 mL of HI 83749-0 Bentocheck reagent to the cuvet. For a correct use of the automatic pipette please follow the related Instruction Sheet. • Replace the cap. Invert several times to mix and then wait for 1 minute. • Insert the cuvet into the instrument and close the lid. • Press READ and “----” will blink on the display. After a few seconds the instrument displays the first reading. The meter continues updating the readings until the “Lamp and Read Status” indicator turns off. e.g. Sample #1 T1