Hanna Instruments HI 8614N User Manual
Page 9

16
17
ELECTRODE CONDITIONING
ELECTRODE CONDITIONING
ELECTRODE CONDITIONING
ELECTRODE CONDITIONING
ELECTRODE CONDITIONING
& MAINTENANCE
& MAINTENANCE
& MAINTENANCE
& MAINTENANCE
& MAINTENANCE
PREPARATION PROCEDURE
Remove the protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT.
This is normal with electrodes and they will disappear when rinsed
with water.
During transport tiny bubbles of air may have formed inside the glass
bulb. The electrode cannot function properly under these conditions.
These bubbles can be removed by "shaking down" the electrode as
you would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in
HI 70300
storage solution for at least one hour.
Please note:
HI 8615 or HI 8615L
-1000 mV
=
4 mA
0 mV
=
12 mA
1000 mV
=
20 mA
350 mV
=
14.8 mA
200 mV
=
13.6 mA
275 mV
=
14.2 mA
1 mV
=
12.008 mA
Note: When the HI 8615 and HI 8615L are used in conjunction
with the Hanna new
HI 8512 indicator or HI 8720 controller,
the calibration can also be performed on the indicator/
controller. In this case slight adjustment can be made on the
indicator/controller even if the whole system calibration is
advised, always starting from the transmitter.