Hanna Instruments HI 8614N User Manual
Page 3

4
5
SPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614L
SPECIFICATIONS OF HI 8614 & HI 8614L
Range
4 to 20 mA / 0.00 to 14.00 pH
Resolution
0.01 mA / 0.01 pH
Accuracy
(@20°C/68°F)
±0.02 mA / ±0.02 pH
Typical EMC Deviation
±0.25 mA / ±0.2 pH
Calibration
Offset: ±2.2 mA / ±2 pH
Slope: ±0.5 mA / 86 to 116%
Temperature Compensation
Fixed or automatic from 0 to 100°C
(32 to 212°F) with
HI 76608 probe
Input Impedance
10
12
Ohm
Output
4-20 mA (isolated)
Installation Category
I I
Protection
IP 65
Environment
0 to 50°C (32 to 122°F);
RH max 95% non-condensing
Power Supply
HI 8614: 18 to 30 Vdc
HI 8614L: 20 to 36 Vdc
Load
Max 500 Ohm
Dimensions
165 x 110 x 71 mm
(6.5 x 4.3 x 2.8'')
Weight
1 kg (2.2 lb.)
The transmitters use a universal BNC socket for quick and secure
connection to any electrode with a BNC connector.
For
HI 8614 family models the temperature compensation is
performed by the transmitter's ATC circuitry if the temperature probe
(
HI 76608, optional) is attached. If ATC is not required, it is possible
to substitute the temperature probe with a fixed resistor.
The input is isolated from the current loop to eliminate problems
related to ground loop, low insulation cables, multiple electrode
connections, and a common mode voltage of up to 100 V ensures true
differential readings.
The terminal board of the transmitter provides for connection of power
supply, pH or ORP electrodes and temperature probe (for
HI 8614
family models only).
The unit is enclosed in a protective casing conforming to IP 65
standards.
Ordering codes:
Code
Range
LCD
HI 8614
0.00 to 14.00 pH
NO
HI 8614L
0.00 to 14.00 pH
YES
HI 8615
±1000 mV
NO
HI 8615L
±1000 mV
YES