SPECIFICATIONS OF HI 8614 & HI 8614L

SPECIFICATIONS OF HI 8614 & HI 8614L

SPECIFICATIONS OF HI 8614 & HI 8614L

SPECIFICATIONS OF HI 8614 & HI 8614L

SPECIFICATIONS OF HI 8614 & HI 8614L

Range

4 to 20 mA / 0.00 to 14.00 pH

Resolution

0.01 mA / 0.01 pH

Accuracy
(@20°C/68°F)

±0.02 mA / ±0.02 pH

Typical EMC Deviation

±0.25 mA / ±0.2 pH

Calibration

Offset: ±2.2 mA / ±2 pH

Slope: ±0.5 mA / 86 to 116%

Temperature Compensation

Fixed or automatic from 0 to 100°C

(32 to 212°F) with

HI 76608 probe

Input Impedance

10

12

Ohm

Output

4-20 mA (isolated)

Installation Category

I I

Protection

IP 65

Environment

0 to 50°C (32 to 122°F);

RH max 95% non-condensing

Power Supply

HI 8614: 18 to 30 Vdc

HI 8614L: 20 to 36 Vdc

Load

Max 500 Ohm

Dimensions

165 x 110 x 71 mm

(6.5 x 4.3 x 2.8'')

Weight

1 kg (2.2 lb.)

The transmitters use a universal BNC socket for quick and secure
connection to any electrode with a BNC connector.

For

HI 8614 family models the temperature compensation is

performed by the transmitter's ATC circuitry if the temperature probe
(

HI 76608, optional) is attached. If ATC is not required, it is possible

to substitute the temperature probe with a fixed resistor.

The input is isolated from the current loop to eliminate problems
related to ground loop, low insulation cables, multiple electrode
connections, and a common mode voltage of up to 100 V ensures true
differential readings.

The terminal board of the transmitter provides for connection of power
supply, pH or ORP electrodes and temperature probe (for

HI 8614

family models only).

The unit is enclosed in a protective casing conforming to IP 65
standards.

Ordering codes:

Code

Range

LCD

HI 8614

0.00 to 14.00 pH

NO

HI 8614L

0.00 to 14.00 pH

YES

HI 8615

±1000 mV

NO

HI 8615L

±1000 mV

YES