The

HI 88703 has built in complete GLP information. The calibration date and the calibration points

are displayed in a comprehensive mode for each range.
To display the GLP information, simply press

CAL key. A screen with instrument serial number and

with information about the calibration is displayed. For further information, press the

GLP key.

The GLP contains:

• Instrument serial number
• The last user calibration date, in selected

format and time in hh.mm format. If no calibration
was performed, the “

Factory Calibration” message

is displayed.

• Parameter as Ratio Turbidity or Non Ratio

Turbidity.

• The value of each calibration point (up to

5 points for ratio turbidimeter and 2 points for non
ratio turbidimeter). If the first calibration point was
skipped, the 0.00 value is displayed.

GOOD LABORATORY PRACTICE (GLP)

To restore the factory calibration for the currently
selected range, press

CAL key when in the main

screen. The first screen for the GLP is displayed.
Press the

Delete key to initiate the calibration delete

procedure and then press the

CFM key to delete the

user calibration and restore the Factory calibration.
Note: Only the user calibration for the current

selected range is deleted.

RESTORE FACTORY CALIBRATION