• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

• Always use this position to align it with the

sign on the instrument top.

In order to match multiple cuvettes follow the next
steps:
• Fill some cuvettes with high quality water

(<0.1NTU) up to the 10 mL mark.

• Clean and oil the cuvettes as described before.

• Turn the instrument ON.

• Insert the first cuvette into the instrument and

press the

Read key. Record the reading.

• Mark this position on the thicker white band on

the top of the cuvette with a water resistant
pencil.

MATCHING MULTIPLE CUVETTES

Precise measurements require the use of a single cuvette. If it is not possible, cuvette selection
and matching must be performed before taking measurements.