Specifications – Hanna Instruments HI 93102 User Manual
Page 7

13
12
HI 93102
Range Turbidity
0.00 to 50.0 NTU*
Bromine
0.00 to 8.00 mg/L
Free Cl
2
0.00 to 2.50 mg/L
Total Cl
2
0.00 to 3.50 mg/L
CYS
0 to 80 mg/L
Iodine
0.0 to 12.5 mg/L
LR Iron
0.00 to 1.00 mg/L
pH
5.9 to 8.5
Resolution
Turbidity
0.01 and 0.1 NTU*
Bromine
0.01 mg/L
Free Cl
2
0.01 mg/L
Total Cl
2
0.01 mg/L
CYS
1 mg/L
Iodine
0.1 mg/L
LR Iron
0.01 mg/L
pH
0.1
Accuracy
Turbidity
±0.5 NTU* or ±5% (whichever greater)
Bromine
±0.08 mg/L; ±3%
Free Cl
2
±0.03 mg/L; ±3%
Total Cl
2
±0.03 mg/L; ±3%
CYS
±1 mg/L; ±15%
Iodine
±0.1 mg/L; ±5%
LR Iron
±0.02 mg/L; ±3%
pH
0.1 pH
Light Source
Pure Green LED
Light Source Life
Life of the instrument
Light Detector
Two Silicon Photocells
Power Source
4 x 1.5V AA alkaline batteries
Battery Life
60 hours or 1000 measurements
Auto-off
Selectable after 10, 20, 30,
40, 50 or 60 min. of non-use
Environment
0 to 50°C (32 to 122°F)
RH max 95% (non-condensing)
Dimensions
220 x 82 x 66 mm
(8.7 x 3.2 x 2.6")
Weight
510 g (1.1 lb.)
SPECIFICATIONS
* 1 NTU = 1 FTU
shown. If pressed in
time/date setup mode, the
meter will quit current
mode without making any
changes to current time/
date.
12) key
Scrolls upwards through
the parameters to be mea-
sured. In calibration/diag-
nostic mode, increments
the blinking digit by one.
If pressed together with
ALT while the meter is in
logging mode, the upper
LCD will show the data
(date/time/value) in the
memory.
13) ALT key:
Alternative functions.
14) key:
Scrolls downwards through
the parameters to be mea-
sured. In calibration/diag-
nostic mode, decreases
the blinking digit by one.
If pressed together with
ALT while the meter is in
logging mode, the upper
LCD will show the current
lot number.