Hanna Instruments HI 93102 User Manual
Page 16
![background image](/manuals/737319/16/background.png)
31
30
LOGGING WITH C 102
HI 93102 allows user to log 25 time/day-tagged
measurements. User can easily turn the log-
ging mode on and off, review the logged
memory, review the current lot number and
clean the buffer (memory). C 102 also re-
minds user if its memory is full.
TURNING THE LOG MODE ON OR OFF
Enter the diagnostic
mode by pressing ALT
and CAL together.
Select mode 12 and press the
key.
The display will show the current (vacant) lot
together with “LOG” if the log mode is on.
Otherwise it will show “----” if the log mode is
off.
By pressing the
key, the meter toggles
between the Log on and off positions. If the
log on mode is selected, every time a mea-
surement is taken (READ pressed) the rel-
evant values will be stored in the current
(vacant) lot number.
To quit diagnostic mode,
press the ALT and CAL
keys together again.
The LCD will then show “diAG” and “Stor” for a
few seconds. If the log on mode was selected
“LOG” will appear on the bottom left hand side
+
+
To quit diagnostic mode,
press the ALT and CAL
keys together again.
+
DIAGNOSTIC MODE
HI 93102 facilitates operations by providing a
diagnostic mode. In this mode, user can set
or verify different parameters necessary to
ensure optimum performance of the meter.
To enter the diagnostic mode, turn the meter
on and momentarily press ALT and CAL to-
gether. The display will show four dashes to-
gether with “d 00”:
Using the , and
keys, select the re-
quired diagnostic mode and press the
key.
The meter will execute one of the following
user-diagnostic functions:
10
Customizes automatic shutdown
11
Selects User or Factory functions
12
Turns logging mode on or off
21
Calibrates span in turbidimetric mode
31
Calibrates span in colorimetric mode
40
Clears the logged memory
The following diagnostic modes are reserved
for authorized service technicians:
00
Shows the
Blank level in colorimetric mode
01
Shows the
Sample level in colorimetric mode
02
Shows the
Dark level in colorimetric mode
05
Shows the
Ground voltage
06
Shows 5V on-board level
07
Shows battery voltage level
08
Shows 1.23V reference voltage level
09
Shows -5V on-board level
99
Shows software version number
+