beautypg.com

Hanna Instruments HI 93102 User Manual

Page 16

background image

31

30

LOGGING WITH C 102

HI 93102 allows user to log 25 time/day-tagged
measurements. User can easily turn the log-
ging mode on and off, review the logged
memory, review the current lot number and
clean the buffer (memory). C 102 also re-
minds user if its memory is full.

TURNING THE LOG MODE ON OR OFF

Enter the diagnostic
mode by pressing ALT
and CAL together.

Select mode 12 and press the
key.

The display will show the current (vacant) lot
together with “LOG” if the log mode is on.
Otherwise it will show “----” if the log mode is
off.

By pressing the

key, the meter toggles

between the Log on and off positions. If the
log on mode is selected, every time a mea-
surement is taken (READ pressed) the rel-
evant values will be stored in the current
(vacant) lot number.

To quit diagnostic mode,
press the ALT and CAL
keys together again.

The LCD will then show “diAG” and “Stor” for a
few seconds. If the log on mode was selected
“LOG” will appear on the bottom left hand side

+

+

To quit diagnostic mode,
press the ALT and CAL
keys together again.

+

DIAGNOSTIC MODE

HI 93102 facilitates operations by providing a
diagnostic mode. In this mode, user can set
or verify different parameters necessary to
ensure optimum performance of the meter.
To enter the diagnostic mode, turn the meter
on and momentarily press ALT and CAL to-
gether. The display will show four dashes to-
gether with “d 00”:

Using the , and

keys, select the re-

quired diagnostic mode and press the

key.

The meter will execute one of the following
user-diagnostic functions:

10

Customizes automatic shutdown

11

Selects User or Factory functions

12

Turns logging mode on or off

21

Calibrates span in turbidimetric mode

31

Calibrates span in colorimetric mode

40

Clears the logged memory

The following diagnostic modes are reserved
for authorized service technicians:

00

Shows the

Blank level in colorimetric mode

01

Shows the

Sample level in colorimetric mode

02

Shows the

Dark level in colorimetric mode

05

Shows the

Ground voltage

06

Shows 5V on-board level

07

Shows battery voltage level

08

Shows 1.23V reference voltage level

09

Shows -5V on-board level

99

Shows software version number

+