A slightly scratched crystal may benefit from a treat-
ment with a fluoride containing toothpaste (without
baking soda). Lightly wipe the sensor tip with the
mild abrasive. Rinse well with deionized and soak in
a fluoride standard near measurement value.

Avoid large changes in temperature (thermal shock)
as it may damage the sensor.

HI 4010

Remove protective cover from sensor tip.

HI 4110

Remove the protective plastic wrap that covers the
ceramic junction before assembling sensor for the first
time.

Ensure o-ring is installed on sensing module before
screwing into the inner stem.

O-Ring

Add reference HI 7075 fill solution daily to maintain
a good head pressure. For optimum response, this
level should be maintained and not be allowed to
drop more than 2-3 cm (1-inch) below fill hole.

During measurement always operate electrode with
the fill hole open.

During normal use, fill solution will slowly drain out
of the tapered cone junction at the lower portion of the
electrode. Excessive loss (>4 cm drop within 24 hours)
is not normal. If this occurs verify cap is tightened and
the interface between the internal cone and outer
body is free of debris.

If an erratic measurement occurs, check to see if for-
eign matter is seen trapped near the internal cone.
Drain and refill with fresh fill solution.