If no fluoride complexing species are in the sample, TISAB
needn’t be added at the same proportions. Add 1 parts
TISAB to 100 parts sample or standard. Ionic strength will
be fixed at approximately 0.02M. Attention: Always pre-
pare samples and standards with the same ISA to volume
ratio.
1) Follow sections VIII and IX to prepare sensors for

measurement.

2) Follow section VI to prepare standards/solution. Stan-

dards should bracket the sample concentration. Stan-
dards and solutions should be at the same tempera-
ture.
1 part of TISAB is added to 100 parts of both samples
and standards. Add stir bar and mix before taking
measurements.

3) Follow section VII; General Guidelines to optimize test

set-up.

4) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable measurement
before recording values. Longer equilibrations are re-
quired at lower concentrations (4-5 minutes).

5) To prevent carry over and contamination of samples,

rinse sensors with deionized water and dab dry be-
tween samples.

Typical Linearity for HI 4010 and HI 4110 Sensors

-150.00

-100.00

-50.00

0.00

50.00

100.00

150.00

200.00

250.00

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

-Log of Concentration

m

V