The Hanna HI 4010 and HI 4110 are ion selective elec-
trodes designed for the measurement of fluoride ions in
aqueous solutions. The HI 4010 is a solid state half-cell
sensor that requires a separate reference. The HI 4110 is a
combination ion selective electrode.
II
IIII
IIII.
Specifications
Specifications
Specifications
Specifications
Specifications
Type:
Solid State electrode with
a Lanthanum Fluoride
crystal membrane.
Ion(s) measured:
Fluoride (F
-
)
Measurement range: Saturated to 1X 10
-6
M
Saturated to 0.02 ppm
Interfering ions:
OH
-
Note:
Several other ions (Al
3+
, Fe
3+
) that complex with the
measured species will reduce the ion concentrations
measured directly. TISAB reagent should be used in most of
these cases. H
+
ion also forms HF species below pH 5.
Increase pH in these cases above 5 for a total fluoride
measurement.
Operating Temperature:
0-80°C
Operating pH:
5 to 8 pH
Dimensions:
12 mm (OD) X 120 mm
nominal insertion
(0.47” X 4.72”)
Connection:
BNC