Hanna Instruments HI 4421 User Manual

Page 49

background image

49

LOGGING MODE 3
This logging mode can be used for any sample measurements. By
choosing this logging mode,

will be available in

Measure

mode.
To log data using this mode:
• Press

while in

Measure

mode to manually log a

record. The “Logged” indicator will be displayed on the LCD.

• The records will be stored in one lot. In order to change the

logging lot, see the measured parameter setup for details, Log
option, New Lot generation.

LOGGING MODE 4
This logging mode can be used for multiple samples measurement. By choosing this logging mode,
and

will be available in

Measure

mode.

To log data using this mode:
• Press

while in

Measure

mode to manually log a record. Each value is logged at the time when the key

was pressed. When the measured value is frozen on the LCD by pressing

and the stability criterion is

reached, the logged value is the one that has been frozen on the LCD.

• To store another frozen value, press

to return to normal logging mode and then

again.

• The records will be stored in one lot. In order to change the logging lot, see the measured unit

Setup

for

details, Log option, New Lot generation.

LOGGING MODE 5
This logging mode can be used for multiple samples measurement. By choosing this logging mode,
and

will be available in

Measure

mode.

Note: If the Reading Mode option is set as Direct and the Logging Mode 5 session is started, a warning pop-
up will be displayed on the LCD, informing the user that the Reading Mode option must be set as Direct/
AutoHold in order to use this logging mode.
To log data using this mode:
• Press

while in

Measure

mode to start the logging session. The logged values are only the ones

frozen on the LCD, after

was pressed and the stability criterion reached.

• To store another frozen value, press

to return to normal logging mode and then

again.