Keep the protective cap on during polarization time and remove it for calibration and measurements.

To calibrate the meter:
• Insert and rinse the probe in the first beaker in order to

decontaminate it;

• Insert the probe in the second beaker;
• Tap the probe repeatedly to remove any air bubbles that may

be trapped inside the sleeve.

• Enter in calibration mode by pressing

;

• Wait to stabilize;
• You can delete a previously performed calibration by pressing

.

When the automatic standard recognition is selected:

• The calibration point will be automatically selected from the two standards available.
• Press

to perform the calibration or

to exit calibration.

Note:If you want to perform automatic DO calibration in two points, perform the calibration at 0% saturation

first and then at 100% saturation.

When the user standard is selected:
• The calibration can be performed only in one point.
• Edit the desired standard value by using

and

.

• Press

to finish the calibration or

to exit calibration.