The OUR is used to determine the oxygen consumption or respiration rate. It is defined as the mg/L of oxygen
consumed per hour.
The following equation is used for OUR determination:

where:
DO

START

= Dissolved oxygen level at start of test

DO

END

= Dissolved oxygen level at end of test

t

ELAPSED

= Elapsed time of test in seconds

total volume/sample volume = Dilution factor of sample

Before starting an OUR measurement remember to set the OUR configuration from the OUR setup menu and make
sure the instrument has been calibrated on DO.

To measure the OUR of a sample:
• Press

and then

to select OUR measure mode.

• Press

to start taking the measurement.

• At the end of the measurement the meter will display the

computed OUR value, the duration of the measurement and
the pressure and temperature values.

OUR MEASURE

OUR MEASURE

OUR MEASURE

OUR MEASURE

OUR MEASUREMENT